Pumps

Rx level electrode (ORP) with cable, glass

Электрод уровня Rx (ОВП) с кабелем, стекло

With 1m cable. or 5 m.

Rx level sensor (ORP) with cable, glass - used in Etatron automatic water treatment stations.
The sensor body with a diameter of 12 mm is made of glass filled with a transport electrolyte in the form of a gel, which guarantees operation at pressures up to 6 bar, supplied with a coaxial cable from 1 to 5 m in length with a BNC connector.

RX sensor specifications:

  • Sensor height: 120 mm.
  • Working range ± 2000 mV.
  • Calibration fluid, type: Rx 475 mV, 650 mV.
  • Maximum working temperature: 80 ° C.
  • Connector type: BNC.
Name vendor code
Rx sensor 5.5 bar - 80 ° C, wire 1 meter AEL0002302
Rx sensor 5.5 bar - 80 ° C, wire 5 meters AEL0002303

Calibration:

Wash and dry the electrode by shaking but not wiping, do not clean the probe, always use new buffers.

• Calibration: according to the application, about once a month.
• Continuous operation period is 1000 hours, after which it must be placed in KCL solution to restore sensitivity. It is recommended to install a filter in front of the flow housing.
• It is not possible to determine the sensor life externally.

Cleaning:

If the readings are unstable, when the sensor reads slowly or intermittently, it is possible to restore normal operation by cleaning if there are deposits on the electrodes. To do this, it is necessary to clean metal surfaces with a material of little abrasion, such as toothpaste.

Storage:

The sensor must not be allowed to dry out; during short-term storage, it may be placed in water.

This category does not contain any products.
Sale
Price:
$7,182.33
Pre-order only
- +
13936_22__CAR/SIC__EPDM__304__G60
Mechanical seal R-MG1 22, CAR/SIC, EPDM, 304, G60
Price:
$39.53
Pre-order only
- +
Price:
$90.21
Pre-order only
- +
Price:
On request
Pre-order only
- +
Price:
On request
Pre-order only
- +
0
Select site language:
deuusapolrus